Level 1 Rapid Infuser
The Level 1 system can warm fluid to a temperature of close to 370
Celsius and infuse fluids at rapid rates. These rates depend on the size of
cannula.
Administration of Crystalloids under pressure (300mmHg) | ||
| Cannula Size | Flow rate (ml/min) | Elapsed Time (min/L) |
| 18g | 240 | 4.2 |
| 16g | 440 | 2.3 |
| 14g | 500 | 1.7 |
Administration of Packed Red Cells under pressure (300mmHg) | ||
| Cannula Size | Flow rate (ml/min) | Elapsed Time (min/L) |
| 18g | 150 | 2.3 |
| 16g | 220 | 1.5 |
| 14g | 300 | 1.1 |
Equipment required |
Procedure |
|
|
![]() |
Infusion of blood products
For blood products, or 500ml bags of IV fluids, use the hook on the inside of the door. Otherwise, repeat as above.
Use of non-collapsible containers (e.g. Gelofusion)
It is essential to remove any air from these containers prior to administration under pressure to prevent possible air embolism by:
- Venting of the container using a ‘Braun aeration needle’
- Venting of the container using a giving set.
| Content by SN Miriam Rigney, Dr Íomhar O'
Sullivan 12/03/2004. Reviewed by Dr ÍOS 18/05/2005.